ASTM F1472
ASTM F1472
Revision: “-20a” issued 2020
Alloy: 6AL-4V (UNS R56400)
Product Forms: Wrought Titanium-6Aluminum-4Vanadium Alloy for Surgical
Implant Applications (Bars, Forgings, Sheet, Strip, Plate)
Composition Analysis | |
---|---|
Aluminum: 5.5 – 6.75% | Nitrogen (Maximum): 0.05% |
Vanadium: 3.5 – 4.5% | Hydrogen (Maximum): 0.015% |
Iron (Maximum): 0.30% | Yttrium (Maximum): 0.005% |
Oxygen (Maximum): 0.20% | Titanium: Balance |
Carbon (Maximum): 0.08% |
Tensile Properties (Minimums)* | |
---|---|
Tensile: 130 – 135 ksi | Elongation: 8 – 10% |
Yield: 120 – 126 ksi | Reduction of Area: 15 – 25% |
*Tensile property requirements vary based on orientation, section thickness/diameter, and cognizant engineering organization requirements.
Required Acceptance Tests
- Chemical Composition
- Tensile Properties
- Microstructure
- Surface Contamination
Disclaimer
The use of this information is strictly voluntary. Information may contain typographical errors, and should be used as a guideline only. This Material Spec Sheet contains generalizations and is in no way a substitution for the ASTM F1472 specification.
To purchase ASTM F1472-20a, please visit: http://www.astm.org/Standards/F1472.htm
Changes may be made to the information contained herein.